YSD16V -vet
> 12" Color TFT Display,led backlight ,800x600 pixels.
> Power supply : AC100~240v 50/60Hz
input current : 1.1 ~ 0.5 A 
safety class : category I 
> Battery
Type: Rechargeable lithium battery,
Operating time :> 2.5 hours 
> Trend & Reviewing 
Trend: long trend 96h,min resolution is 1 mins(store when power goes off )
High resolution trend: 2h, mini resolution is 5


NIBP measurement reviewing:400 groups 
Standard parameters
3/5-Lead ECG,Hwatime SpO2, NIBP, RESP, 2-Temp,PR
HR
Range : 15 ~ 350 bpm
Resolution : 1bpm 
Accuracy : ± 1% or± 1 bpm,
whichever is greater
ST segment
Range : -2.0mV to + 2.0mV 15 ~ 350 bpm
Resolution : 0.01mV 
Accuracy : 
-0.8mV to +0.8mV ,± 0.02mV or± 10% ,
over ±0.8mV unspecifie 
which is greater
over ±0.8Mv unspecified 
Respiration Rate
| Measurement method | Thoracic impedance | 
| Lead selec from | II (RA-LL )/default II | 
| Gain | x0.25,x1,x2,x4 | 
| Bandwidth | 0.25Hz to 2.0 Hz(-3dB) | 
| Sweep speed | 6.25mm/s,12.5mm/s,25mm/s | 
| Resolution | 1rpm | 
| Accuracy | ± 2 rpm | 
| Delay of apnea alarm | 10s,15s,20s,25s,30s,35,40s | 
NIBP
| Measurement way | Automatic Oscillometry | 
| Range | Horse 40~ 270 mmHg | 
| Dog 40~ 200mmHg | |
| Cat 40~ 135mmHg | |
| Cuff pressure range | 0mmHg to 300 mmHg | 
| Resolution | 1 mmHg | 
Pressure accuracy
| Static | 3mmHg | 
| Clinic average error | 5mmHg | 
| Standard deviation | =8mmHg | 
| Unit mmHg, kPa | |