Features:
15" color TFT screen Multi-channel display Big font interface YSD18D
ST and arrhythmia analysis
Drug dose calculation and titration table
OxyCRG dynamic view display
Respiration monitoring with apnea alarms
Alarm review, arrhythmia review
Patient info. Input
Anti-defibrillation design
NIBP over pressure protection
Plug-in rechargeable battery, AC/DC available. Supporting wire or wireless network.
ECG
Lead mode               5 Leads, I, II, III, AVR, AVL, AVF, V
Gain                   ×2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV,auto
Heart rate             Measure range:
                        Adult    15 ~ 300 bpm
                        Neonatal/Pediatric 15 ~ 350 bpm
Accuracy                ±1%
Resolution              1 bpm
Sensitivity             > 200 VμP-P
Differential Input Impedance      > 5 M ohm
CMRR                        Monitor Mode   >100 dB
                               Surgery Mode >100 dB
                               Diagnostic Mode  >90 dB
Electrode offset potential     ±300mV
Leakage Current                < 10 μA
PACE pulse detect              range±2~±700mV
                               width 0.1~2ms
                               rise time 10~100 μs
PACE pulse rejection           range ±2 ~±700mV
                               width 0.1~2ms
                               rise time 10 ~100 μs
Baseline Recovery              < 3 s  After defibrillation.
Signal Range                   ±8 mV p-p
Bandwidth        Surgery      1 ~ 20 Hz
                               Monitor 0.5 ~ 40 Hz
                               Diagnostic   0.05 ~ 130 Hz
Calibration Signal             1 mV p-p, 5% accuracy
ST measurement                 Range -2.0 ~+2.0 mV
                               Accuracy-0.8mV~+0.8mV
                               ±0.02mV or±10%, which is greater
                               Other range: unspecified
Respiration
Method                         Impedance between RA-LL
Differential Input Impedance   >2.5 MΩ
Respiration Impedance Range     0.3~3Ω
Base Impedance Range           200Ω~4000Ω
Bandwidth                     0.1 ~ 2.5 Hz
Gain                          ×0.25,×0.500,×1,×2,×4
Respiration Rate               Measurement Range
                               Adult 0 ~ 120 BrPM
                               Neonatal / Pediatric 0 ~ 150 BrPM
Resolution                     1 BrPM
Accuracy                0~6 BrPM: unspecified
                      7~150 BrPM:±2 BrPM or±2%
                      use the greater
Inspiriting current        < 300     A RMS max
Apnea Alarm                10 ~ 40 s
NIBP
Method                            Oscillometry
Measure mode                      Manual, Auto, Continual
Measure Interval in AUTO Mode     1,2,3,4,5,10,15,30,60,90,120,180,240,480 min
Measure Period in Continual Mode    5 min
Pulse Rate Range               40 ~ 240 bpm
Measure and Alarm Range        Adult Mode
                                 SYS 40 ~ 270 mmHg
                                 DIA 10 ~ 215 mmHg
                                 MEAN20 ~ 235 mmHg
                               Pediatric Mode
                                 SYS 40 ~ 220 mmHg
                                 DIA 10 ~ 150 mmHg
                                 MEAN20 ~ 165 mmHg
                               Neonatal Mode
                                 SYS 40 ~ 135 mmHg
                                 DIA 10 ~ 100 mmHg
                                 MEAN 20 ~ 110 mmHg
Static pressure accuracy       ±3mmHg
Resolution                     1mmHg
Accuracy                       Maximum Mean error  ±5mmHg
                               Maximum Standard deviation     8mmHg
Overpressure Protection        Adult 297±3 mmHg
                               Pediatric 240±3 mmHg
                               Neonatal 147±3 mmHg
SpO2
Measurement Range              0 ~ 100 %
Resolution                     1 %
Accuracy                      70% ~ 100% ±2 %
                              69%
                              0% ~ 39%  unspecified
Pulse Rate                    Measure and Alarm Range 20~300bpm
                              Resolution 1bpm
                              Accuracy ±3bpm
TEMP
Channel                       2
Measure and Alarm Range       0 ~ 50ºC
Resolution                    0.1ºC
Accuracy(no sensor)           ±0.1(25ºC-45ºC),0.2±ºC(other)
Accuracy(include sensor)      ±0.2ºC(32ºC-42ºC)
                     ±0.3(other)
tandard:  ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional:  EtCO2, IBP, Thermal Recorder

Why choose us:
we got production and business certification from Chengdu food and Drug Administration
we got business license from Chengdu City Administration of industry and commerce
we got products certificate of registration and have checked in the website of SFDA
we have accepted the onsite check from SFDA every quarter
we finished the training of laws and regulations in medical instrument field
We have well-established system about material purchasing, employee training,store management,after-sales service process.etc
We adheres the beliefs of "Support Healthy Life", and our company spirits are quality, service,trust,thanksgiving;